1. A Novel Polymeric Membrane Sensor for Chlorhexidine Determination.
- Author
-
Lenik J and Sokal K
- Subjects
- Electrodes, Potentiometry, Polymers, Hydrogen-Ion Concentration, Chlorhexidine, Membranes, Artificial
- Abstract
In the present work, potentiometric sensors with polymer membranes used for chlorhexidine (CHXD) determination were developed. The polymer membranes were plasticized with bis(2-ethylheksyl)sebacate (DOS) or 2-nitrophenyloctyl ether ( o -NPOE). The active compounds used in the membrane were cyclodextrins, crown ethers, and ion exchangers. The best-constructed electrode was based on neutral heptakis(2,3,6-tri-O-benzoyl)- β -cyclodextrin with lipophilic salt (KTpClBP)-potassium tetrakis(4-chlorophenyl) borate-dissolved in plasticizer, DOS. The presented electrode is characterized by an average cationic slope of 30.9 ± 2.9 mV decade
-1 within a linear range of 1 × 10-6 to 1 × 10-3 mol × L-1 , while the value of the correlation coefficient is 0.9970 ± 0.0026. The response time was about 5 s when increasing the sample concentration and about 10 s when diluting the sample. The electrode potential is independent of the pH within a range of 4.0-9.5. The polymeric membrane sensor was successfully applied for assays of chlorhexidine digluconate in pure samples and pharmaceutical samples. The relative error from three replicate measurements was determined to be 1.1%. and the accuracy was RSD = 0.3-1.1%.- Published
- 2023
- Full Text
- View/download PDF