1. Comparison between a new computer program and the reference software for gray-scale median analysis of atherosclerotic carotid plaques.
- Author
-
Casella IB, Fukushima RB, Marques AB, Cury MV, and Presti C
- Subjects
- Carotid Artery, Internal diagnostic imaging, Humans, Reproducibility of Results, Software, Carotid Stenosis diagnostic imaging, Plaque, Atherosclerotic diagnostic imaging, Ultrasonography, Doppler, Duplex
- Abstract
Purpose: To compare a new dedicated software program and Adobe Photoshop for gray-scale median (GSM) analysis of B-mode images of carotid plaques., Methods: A series of 42 carotid plaques generating ≥50% diameter stenosis was evaluated by a single observer. The best segment for visualization of internal carotid artery plaque was identified on a single longitudinal view and images were recorded in JPEG format. Plaque analysis was performed by both programs. After normalization of image intensity (blood = 0, adventitial layer = 190), histograms were obtained after manual delineation of plaque. Results were compared with nonparametric Wilcoxon signed rank test and Kendall tau-b correlation analysis., Results: GSM ranged from 00 to 100 with Adobe Photoshop and from 00 to 96 with IMTPC, with a high grade of similarity between image pairs, and a highly significant correlation (R = 0.94, p < .0001)., Conclusions: IMTPC software appears suitable for the GSM analysis of carotid plaques., (© 2014 Wiley Periodicals, Inc.)
- Published
- 2015
- Full Text
- View/download PDF